Confirm oxygen source pressure is within operating range before opening the flowmeter
ContextFor E-cylinders, a full tank reads approximately 2,000 psi; anything below 200 psi should be replaced before induction begins. Central pipeline systems typically deliver at 50–55 psi at the wall outlet — check the line pressure gauge on the machine to confirm it falls in the green zone. If using a pipeline and the gauge reads zero, verify the zone valve has not been accidentally closed and switch to a backup cylinder immediately. Never begin a procedure with only one E-cylinder and no backup if the case is expected to run longer than 45 minutes.